Customization: | Available |
---|---|
Type: | Intelligent Wet and Dry Laser ParticleSize Analyzr |
Sample Feeding: | Dry Dispersion |
Still deciding? Get samples of US$ 18590/Unit
Order Sample
|
Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
---|
Payment Methods: |
|
---|---|
Support payments in USD |
Secure payments: | Every payment you make on Made-in-China.com is protected by the platform. |
---|
Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
This intelligent wet and dry laser particle size analyzer is widely used in cement, ceramics, pharmaceuticals, emulsions, paints, dyestuffs, pigments, fillers, chemicals, catalysts, drilling mud, abrasives, lubricants, coal, silt, dust, cells, bacteria, food additives, pesticides, explosives, graphite, photographic materials, fuels, printing ink, metal and non-metallic powders, calcium carbonate, kaolin, coal slurry, and other powdery materials.
Particle size analyzer adopt laser diffraction theory (Mie and Fraunhofer diffraction), measure size is from 0.01μm to 2000μm (dry 0.1μm-2000μm), Which offer reliable and repeatable particle size analysis for a diverse range of applications.
It use dual-beam & multiple spectral detection systems and side light scatter test technology to significantly improve precision and performance of test.
Intelligent full automatic operation system and manual operation mode, freely choose.
Full built-in sample dispersion system.
Instrument particle size analysis software.
Model | PSA-2308A | |
Principle | Laser diffraction principle | |
Analysis | Mie and Fraunhofer scattering | |
Detector Arrangement | Log-spaced array, test angle from 0.015 degree to 145 degree | |
Measuring Range | Wet: 0.01μm-2000 μm; Dry: 0.1μm-2000μm | |
Silicon Photodetectors | Wet:127 pcs; Dry:100 pcs | |
Accuracy Error | Wet<1% Dry<1% (CRM D50) | |
Repeatability Error | Wet<1% Dry<1% (CRM D50) | |
Light Source | High performance semiconductor red laser (λ=639nm ) P>3.0MW | |
Auxiliary green solid semiconductor laser (λ= 405 nm) P>2.0MW (available) | ||
Optical Path | Converging light Fourier transform optical path | |
Effective Focal Length | 500mm | |
Laser Safety | Class 1 | |
Wet Dispersion | Ultrasonic | Frequency: 40KHz |
Power: 60W | ||
Time: ≥1S | ||
Stir | Revolutions Speed: 0-3000RPM (Adjustable) | |
Circulate | Rated Flow: 30L/min | |
Rated Power: 70W | ||
Water Level Sensor (UK) | Prevent water overflow and effectively protect the instrument | |
Sample Tank | Volume: 1000mL | |
Micro-Sample Cuvette | Volume: 10mL ( Available) | |
Dry Dispersion | Dry-turbulence dispersion patent technology, normal shock wave shear technique | |
Feeding Speed | Adjustable (Variable speed knob) | |
Operation Mode | Full automatic / manual control, freely choose | |
Dispersion Medium | Compressed Air, pressure: 0 to 6 bar | |
Optical Bench Alignment System | Full automatic, precision is up to 0.2um | |
Full Test Speed Per Time | Wet: <2 min | |
Dry : <1 min | ||
Inverval time per test result: 500ms | ||
External Size(L×W×H)(mm) | 1040×440×540 | |
Shipping Size (W×D×H)(mm) | 1220x640x700 | |
N.W. / G.W. (kg) | 70/99 |