Customization: | Available |
---|---|
After-sales Service: | 24-hour online service,free accessories |
Application: | Laboratory Apparatus |
Still deciding? Get samples of US$ 3510/unit
Order Sample
|
Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
---|
Payment Methods: |
|
---|---|
Support payments in USD |
Secure payments: | Every payment you make at Made-in-china.com is protected by the platform. |
---|
Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Reliable cold light source, with low power consumption, long life and other advantages.
High detection sensitivity is suitable for analysis and measurement of ultra-low concentration samples.
Fluorescence intensity mode allows fluorescence spectrum scanning, fluorescence kinetic determination and quantitative analysis.
Provide ten emission spectrum scanning speed selection, can complete the full spectrum scanning within 1 second.
Model | SP-FL96 |
---|---|
Excitation light source | High brightness LED |
Excitation LED | 250nm ~ 600nm(LED) |
EX wavelength standard set: 365nm,376nm,392nm,405nm | |
Emission monochromator |
C-T configuration diffraction grating monochromator Emission wavelength range (EM): 200nm-650nm, Bandwidth: 10nm |
(Extend the monochromator to Em200-900 is optional.) | |
Emission wavelength accuracy: ± 1nm | |
Emission wavelength reproducibility: ≤ 0.5nm | |
S/N ratio | S/N≥90(Using 1cm quartz sample cell, measure the signal |
noise ratio of Raman spectrum of water ) | |
Detection limit | 1×10-10 g/ml quinine sulfate solution |
Linearity (γ) | ≥0.995 |
Repetitive peak intensity | ≤1.5% |
Zero drift | ≤0.3(within 10min) |
Upper limit change of indicating value | ≤1.5% (Within 10 minutes) (displaying value≥50) |
Power type | 220V±22V; 110V±22V |
Size(W×D×H)(mm) | 442×392×250 |
Shipping size(W×D×H)(mm) | 535x500x380 |
N.W./G.W. (kg) | 10/15 |