Customization: | Available |
---|---|
Theory: | Autoclave |
Type: | Pressure Steam Sterilization Equipments |
Still deciding? Get samples of US$ 39100/Unit
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Translation door, with anti-pinch hand function. Vacuum sterilizer also adopts a double-door channel design, which is not only equipped with a powerful pressure safety interlock function but also a temperature safety interlock on the liquid process door to guarantee the safety of operation in all aspects.
The high-efficiency design of the automatic sealing ring makes its sealing effect outstanding. The service life is up to 1500 times, providing users with long-lasting and stable performance.
The main structure is european ring reinforcement jacket structure, programmable controller PLC, color touch screen.
Provides a wealth of customized procedures, covering aseptic clothing, utensils, rubber, liquids, media, waste, sterilization, and many other areas, but also includes BD test, vacuum leakage test, and other professional test procedures to meet the needs of different areas.
Vacuum autoclave has a built-in micro-needle printer and a paperless recorder to make recording and tracking more convenient.
Automatic drainage device, optional jacket automatic control drainage device, the inner chamber are installed automatic control drainage device, automatic detection jacket and inner chamber temperature pressure automatic drainage, to ensure the saturation of steam.
Intelligent maintenance system: Optional remote monitoring and maintenance module; Water saving noise reduction system is optional
With drainage temperature automatic adjustment system, drainage temperature ≤55°C.
The ST-BVH series pulse vacuum autoclave utilizes saturated steam as the medium, employing pulsating vacuum for the efficient eradication of microorganisms and their spores - marking it as the most effective method for physical sterilization. The adoption of pulsating vacuum exhaust serves to eliminate the impact of cold air on temperature, ultimately achieving sterilization through the combined processes of vacuum dehumidification and jacket drying.
Pulse vacuum autoclave is specifically tailored to meet the high demands of various industries, including pharmaceutical biological engineering, medical health, and laboratory animal research. It serves as an indispensable solution for the sterilization of equipment, aseptic clothing, rubber stoppers, aluminum caps, raw materials, filters, media, and other items with stringent sterilization requirements.
The saturated steam, in tandem with the pulsating vacuum technology, ensures an efficient and thorough sterilization process, effectively eliminating microorganisms and their resilient spores. By incorporating pulsating vacuum exhaust, the sterilizer further guarantees the elimination of cold air influences on temperature, ensuring optimal conditions for the sterilization of a diverse range of items. The ST-BVH series pulse vacuum autoclave stands out as a reliable and versatile tool for achieving precise and effective sterilization in critical applications across various fields.
Model | ST-BVH1200PV | ST-BVH1500PV | ST-BVH2000PV | ST-BVH2500PV | ST-BVH3000PV | ST-BVH6000PV | ST-BVH6700PV | ST-BVH13600PV |
---|---|---|---|---|---|---|---|---|
Capacity(L) | 1200 | 1500 | 2000 | 2500 | 3000 | 6000 | 6700 | 13600 |
Interior Design Pressure | -0.1~03MPa | |||||||
Jacket Design Pressure | 0.3MPa | |||||||
Design Temperature | 150ºC | |||||||
Max Working Temperature |
139ºC | |||||||
Max Working Pressure | 0.25MPa | |||||||
Vacuum Leakage Rate | ≤0.13KPa/min | |||||||
Final Vacuum | -96KPa | |||||||
Opening Pressure of the Inner Chamber Safety Valve |
0.28MPa | |||||||
Jacket Safety Valve Opening Pressure |
0.28MPa | |||||||
Sterilization Temperature Control Precision |
0-2ºC | |||||||
Temperature Uniformity | ±0.5ºC | |||||||
Negative Pressure Pulsating Vacuum Negative Value and Number |
Amplitude -80~0KPa, frequency 1~99 | |||||||
Number of Pulsations Across Pressure |
Amplitude -80 ~80KPa, frequency 1 | |||||||
Number of Positive Pressure Pulsations |
1~3 | |||||||
Inner Tank Size (WxH) (mm) | 680x1180 | 680x1180 | 900x1380 | 900x1380 | 900x1380 | 900x1380 | 1250x1380 | 1250x1380 |
Depth | 1500 | 1800 | 1620 | 2020 | 2400 | 4800 | 3910 | 7850 |
Overall Size (LxWxH) (mm) |
2007x2710 x2173 |
2357x2110 x2713 |
1950x2750 x2200 |
2300x2750 x2200 |
2700x2750 x2200 |
5400x2750 x2200 |
5400x3200 x2200 |
8500x3200 x2200 |
Weight (kg) | 2300 | 2450 | 3000 | 3600 | 3800 | 6600 | 7000 | 11000 |
Gas Consumption (Kg/c) | 65 | 80 | 100 | 120 | 150 | 320 | 350 | 620 |
Water Consumption(Kg/c) | 600 | 750 | 900 | 1100 | 1300 | 2400 | 2600 | 4800 |
Power Supply | 380V,50Hz | 380V,50Hz | 380V,50Hz | 380V,50Hz | 380V,50Hz | 380V,50Hz | 380V,50Hz | 380V,50Hz |
Consumption | 3.5kw+(73kw) | 3.5kw+(90kw) | 4.5kw | 6kw | 8kw | 12kw | 12kw | 24kw |